Statistical calculations
Statistical calculations are carried out only on total mass of prepared formulation (mass of
each ingredient in a formulation is not included in the statistics). If a user wants to run
statistics if the formula making mode, then:
1. Go to parameter 06 Statistics
2. Erase results of previous statistical calculations
3. Set parameter 06 Statistics to YES
4. Enter the formula making mode
5. Carry out required series of measurements
6. Once again go to parameter 06 Statistics
7. Enter parameter 06 02 Results
8. In order to print statistics press PRINT key.
18.8. Statistics
Mode activating
Fig. 54. Statistics mode activating
The very first process after entering the Statistics mode should be erasing the results of
previous statistical calculations. It is carried out by using option M8 01 Erase.
All statistical data are updated on an ongoing basis after saving a measurement result in
balance memory. Adding another measurement to a series is carried out on placing a
weighed load on balance weighing pan, stabilization of measurement result (the measuring
unit appears on the display) and pressing ENTER key.
The user determines whether statistical data should be visible on the graphic of a balance
display while carrying out the measuring process. This parameter is set in statistics mode
submenu (the data is visible on setting the parameter to “YES”).