- 87 -
value in a current
measuring unit
Switch on continuous transmission in a basic measuring unit
C1_I CR LF
- command understood but not
accessible at this moment
- command understood and in progress
MASS FRAME
- response: mass value in a basic
measuring unit
S I space
space character mass space unit CR LF
Switch off continuous transmission in a basic measuring unit
C0_I CR LF
- command understood but not
accessible at this moment
- command understood and carried out
Switch on continuous transmission in a current measuring unit
CU1_I CR LF
- command understood but not
accessible at this moment
- command understood and in progress
MASS FRAME
- response: mass value in a current
measuring unit
S U I
space character mass space unit CR LF
Switch off continuous transmission in a current measuring unit
CU0_I CR LF
- command understood but not
accessible at this moment
- command understood and carried out
Set min checkweighing threshold
Format: DH_XXXXX CR LF, where: _ - space, XXXXX – mass format
ES CR LF
- command not recognised (mass
format incorrect)
Set max checkweighing threshold
Format: UH_XXXXX CR LF, where: _ - space, XXXXX - mass format
ES CR LF
- command not recognised (mass
format incorrect)
Give value of min checkweighing threshold
Response: DH_MASA CR LF - command carried out