- 91 -
number
OK
CR LF
comman
d carried
out
Set working mode
Command overview: Command sets particular working mode.
Format: OMS_n CR LF
Response options:
OMS_OK CR LF
command carried out
OMS_E CR LF
error in-course of command
execution, no parameter or
incorrect format
OMS_I CR LF
command understood but not
accessible at this moment
n parameter, decimal value determining working mode number. To see detailed description go to
OMI command
An example:
Command:
OMS_1
3 CR LF
set
Statistics
mode
Response:
OMS_OK CR
LF
Statistics
mode set
Give current working mode
Command overview: Command returns currently set working mode.
Format: OMG CR LF
Response options:
OMG_n_OK CR LF
command carried out, response:
current working mode
OMG_I CR LF
command understood but not
accessible at this moment
n parameter, decimal value determining working mode number. To see detailed description go to
OMI command.
An example:
Command:
OMG CR
LF
return current
working mode
Response:
OMG_13_OK CR
LF
balance
operates in
Statistics working
mode
Give accessible units
Command overview: Command returns units available for a particular device and for a current
working mode.
Format: UI CR LF
Response options:
UI_”x
1
,x
2
, … x
n
”_OK CR LF
command carried
out, returns
accessible units
UI_I CR LF
command
understood but not
accessible at this
moment
x unit symbols, separated by means of comas
x → g, mg, ct, lb, oz, ozt, dwt, tlh, tls, tlt, tlc, mom, gr, ti, N, baht, tola, u1, u2
An example:
Command:
UI CR LF
return

Table of Contents