EasyManuals Logo

RADWAG R Series User Manual

RADWAG R Series
99 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #90 background imageLoading...
Page #90 background image
- 90 -
OMI CR LF
n_”Nazwa modu” CR LF:
n_”Nazwa modu” CR LF
OK CR LF
command carried out,
response:
accessible working modes
OMI_I CR LF
command understood but not
accessible at this moment
Mode name parameter, working mode name, inserted in between inverted comas. The name takes
form given on a particular balance display, it is provided in a currently selected language.
n parameter, decimal value determining working mode number.
n →
1 Weighing
2
Parts Counting
3
Deviations
4
Dosing
5
Formulas
6
Animal Weighing
8
Solids Density
9
Liquids Density
10
Peak Hold
11
Totalizing
12
Checkweighing
13 Statistics
CAUTION! Working modes numbering is identical for each kind of balance. The numbers are assigned to
working modes names. Some balances give only the number as a response.
An example 1:
Command:
OMI CR LF
return
accessibl
e
working
modes
Response:
OMI CR LF
2_”
Parts counting” CR LF
4_”
Dosing” CR LF
12_”
Checkweighing CR
LF
accessibl
e
working
modes
are given
in return:
mode
number
+ name
OK CR LF
comman
d carried
out
An example 2:
Command:
OMI CR
LF
return
accessibl
e
working
modes
Response:
OMI CR LF
2 CR
LF
4
CR LF
12
CR LF
accessibl
e
working
modes
are given
in return:
mode

Table of Contents