EasyManuals Logo

RADWAG R Series User Manual

RADWAG R Series
99 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #89 background imageLoading...
Page #89 background image
- 89 -
IC_A CR LF
- command understood and in progress
IC_D CR LF
- calibration completed
IC_A CR LF
- command understood and in progress
IC_E CR LF
- time limit exceeded while waiting for a
stable measurement result
IC_I CR LF
- command understood but not
accessible at this moment
Disable automatic internal adjustment of the balance
Format: IC1 CR LF
Response options:
IC1_I CR LF
- command understood but not
accessible at this moment
IC1_E CR LF
- operation disabled, e.g. for verified
balance
IC1_OK CR LF
- command carried out
Operation is disabled for verified balances.
For non-
verified balances the command inhibits internal calibration until it is enabled via IC0
command or until the balance is turned off. The command doe
s not modify settings specifying
calibration start.
Enable automatic internal adjustment of the balance
Format: IC0 CR LF
Response options:
IC0_I CR LF
- command understood but not
accessible at this moment
IC0_OK CR LF
- command carried out
Operation is disabled for verified balances.
Give balance serial number
Format: NB CR LF
Response options:
NB_A_”x” CR LF
- command understood, response:
serial number
NB_I CR LF
- command understood but not
accessible at this moment
x serial number of the device (inserted in between inverted commas)
An example:
Command:
NB CR LF return serial number
Response:
NB_A_”1234567” serial number of the
device ”1234567”
Lock balance keypad
Format: K1 CR LF
Response options:
K1_I CR LF
- command understood but not
accessible at this moment
K1_OK CR LF
- command carried out
Command locks the balance keypad (proximity sensors, touch panel) until the moment of turning the
balance off or until sending K0 command
Unlock balance keypad
Format: K0 CR LF
Response options:
K0_I CR LF
- command understood but not
accessible at this moment
K0_OK CR LF
- command carried out
Give available working modes
Command overview: Command returns accessible working modes.
Format: OMI CR LF
Response options: