- 93 -
CAUTION!
BEEP sound activated via BP command is inhibited if in-course of its activation the sound is activated by
means of other device: keypad, touch panel, proximity sensors.
Send all implemented commands
Format: PC CR LF
Command:
PC CR LF
- send all
implemented
commands
Response:
PC_A_”Z,T,S,SI…”
- command
carried out, the
terminal
displays all
implemented
commands.
Give balance type
Format: BN CR LF
Response options:
BN_A_”x” CR LF
- command understood, response:
balance type
BN_I CR LF
- command understood but not
accessible at this moment
x series of types for a particular balance (in between inverted commas), with general balance type
in front
An example:
Command:
BN CR LF
return balance type
Response:
BN_A_”AS”
balance type: ”AS
R”
Give max capacity
Format: FS CR LF
Response options:
FS_A_”x” CR LF
- command understood, response: Max
capacity
FS_I CR LF
- command understood but not
accessible at this moment
xMax value of reading units (in between inverted commas)
An example:
Command:
FS CR LF
return Max capacity
Response:
FS_A_”220.000
0”
Max capacity: ”220
g”
Give program version
Format: RV CR LF
Response options:
RV_A_”x” CR LF
- command understood, response:
program version
RV_I CR LF
- command understood but not
accessible at this moment
xprogram version (in between inverted commas)
An example:
Command:
RV
CR LF
return program
version
Response:
RV_A_” 1.1.1”
program version:
”1.1.1”
Set AUTOZERO function
Format: A_n CR LF
Response options:
A_OK CR LF
command carried out