11.6.2. Temporary Unit
Temporary unit runs from the moment it is set to the scale shut-down and
restart.
Procedure:
Enter <P9.Unit / 9.2.Unin> submenu.
Press
key, available units are displayed successively one by one.
Options in case when the main unit is [kg]:
kg (kilogram),
g (gram),
lb (pound)*,
N (Newton).
*) unit disabled for verified weighing devices.
Options in case when the main unit is [g]:
g (gram),
kg (kilogram),
ct (carat),
lb (pound)*.
*) unit disabled for verified weighing devices.
Select temporary unit and press key, next go back to the home
screen.
12. ADJUSTMENT
In order to ensure the highest weighing accuracy, it is recommended
to periodically introduce corrective factor of indications to scale memory,
the said factor must be referred to a reference weight. This is so called
adjustment. Adjustment has to be carried out:
prior weighing,
if long breaks between successive measuring series occur,
if the ambient temperature has changed dynamically,
if the scale has been relocated.