- 42 -
17.11.2. Selecting Tare Value in Balance Memory
Enter <tArE> submenu,
Window with the name of first tare <tArE 0> is displayed in tare database
(press
key to select other record):
Press key to use selected tare,
The value of used tare with minus sign and Net pictogram are displayed
in the left upper part of the display:
Caution:
Entered tare value is not saved on balance restart.
18. ADJUSTMENT
option available for non-verified balances exclusively
In order to ensure the highest weighing accuracy, it is recommended to
periodically introduce a corrective factor of indications to balance memory, the
said factor must be referred to the reference mass. Adjustment has to be
carried out prior first weighing or if the ambient temperature has changed
dynamically. Prior adjustment unload the weighing pan.
18.1. Adjustment
Procedure:
Enter <P6.CAL> submenu. Next:

Table of Contents

Other manuals for RADWAG WTC 600

Questions and Answers:

RADWAG WTC 600 Specifications

General IconGeneral
Capacity600 g
Readability0.01 g
CalibrationExternal
DisplayLCD (backlit)
Linearity±0.02 g
Stabilization Time3 s
InterfaceRS 232
Pan Size128 mm
Power SupplyAC adapter
Operating Temperature15 - 30 °C
Repeatability0.01 g