EasyManua.ls Logo

Rae MultiRAE Plus - MultiRAE Plus Start-up: Check Sensor Zero

Default Icon
66 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
The Hazardous Environment Detection Company
MultiRAE Plus Start-up: Check Sensor Zero
TOX1 VOC TOX2
000
LEL
0-3 20.6-21.2
OXY
If any reading is outside these limits,
wait a few minutes then check again.
If readings are in limits use the unit
If readings are still outside these limits
perform “Fresh air calibration”