Rae MultiRAE - User Manual

Rae MultiRAE
93 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
MultiRAE
User’s Guide
Rev D
December
2013
P/N: M01-4003-000

Table of Contents

Other manuals for Rae MultiRAE

Questions and Answers

  • E
    Eric AdamsJul 27, 2025
    What to do if Rae MultiRAE Measuring Instruments shows a 'Pump failed message'?
    • J
      John HernandezJul 27, 2025
      If your Rae Measuring Instruments displays a 'Pump failed message', several issues could be the cause. It might be due to a blocked inlet probe. Clear any obstructions and press the [Y/+] key to reset the pump alarm. Alternatively, direct connection to a gas outlet while the gas value is turned off can trigger this message, so remove the blocking objects and then press [Y/+] key to reset the pump alarm. Also, a contaminated or overly dirty water trap filter can cause the problem, in this case replace the water trap filter. Water condensation along the inlet probe can also be the reason, so be careful not to allow water condensation inside the unit. If none of these are the reason, the pump or pump circuit may be faulty, requiring a pump replacement.
  • J
    Jamie HerreraAug 5, 2025
    Why does my Rae Measuring Instruments display a 'Lamp' message when I turn it on?
    • P
      Pamela MillerAug 5, 2025
      The 'Lamp' message on your Rae Measuring Instruments typically appears due to low ion concentration inside the PID lamp, especially when first powered on in a cold environment. Try turning the unit off and then back on. If the problem persists, the UV lamp or its circuit may be defective, requiring a UV lamp replacement.
  • R
    ronald57Aug 16, 2025
    How to fix Rae Measuring Instruments that won't turn on after charging?
    • B
      Bryan DuffyAug 16, 2025
      If your Rae Measuring Instruments fails to power on after charging, it could be due to a defective charging circuit or a defective battery. Try another charge of battery, or replace the battery or charger.

Summary

Safety Warnings and Cautions

Read Before Operating

Manual must be read by all users for proper operation and maintenance.

Caution: Hazardous Operations

Warnings about operating the monitor with cover removed or in hazardous areas.

Special Conditions for Safe Use

Safe Use Precautions

Restrictions on battery types, charging locations, and electrostatic discharge precautions.

Intrinsic Safety and Marking

Information on certification, intrinsic safety, and product marking details.

Operational Safety and Disposal

Hazardous Area Classifications and Sensor Warnings

Zone/division classifications and sensor installation warnings.

Key Features

All-in-One Monitoring

Highlights multi-threat monitoring for up to six threats, customizable sensors, and wireless access.

Battery Management

Powering the MultiRAE On and Off

Calibration Procedures

Fresh Air and Zero Calibration

Procedures for fresh air, zero, and single-sensor zero calibration.

Span and Bump Test Procedures

Steps for single-sensor span, multi-sensor span, and bump tests.

Instrument Configuration and Settings

Alarm Configuration

Setting alarm limits, modes, and output behaviors.

Datalogging Configuration Options

Setting datalog interval and choosing sensors for logging.

Policy Enforcement Management

Setting Policy Enforcement

Configuring mandatory calibration or bump tests via PC software.

Calibration and Testing Procedures

Manual Alarms Test and Bump Testing

Performing manual alarm tests and daily bump tests.

Zero and Fresh Air Calibration

Important notes and specific procedures for zeroing CO2 and ppb PID sensors.

Span Calibration Methods

Single-sensor, multi-sensor, and 3-point span calibration techniques.

MultiRAE Pro: Multi-Threat and Gamma Operation

Data Transfer and Firmware Management

PC-Based Data Transfer and Configuration

Downloading data, configuring settings, and updating firmware via PC.

AutoRAE 2 Data Transfer

Procedures for transferring data and managing settings via the AutoRAE 2.

Instrument Maintenance

Sensor Module Replacement

Steps for removing, cleaning, and replacing sensor modules.

PID Sensor and Pump Replacement

Procedures for PID sensor maintenance and pump replacement.

Alarms and Notifications

Alarm Signal Summaries

Detailed tables of alarm types and indicators in Hygiene and Search modes.

Troubleshooting Common Issues

Power, Password, and Alarm Issues

Solutions for power, password, and inoperative alarm problems.

Pump and Lamp Troubleshooting

Resolving pump failure messages and PID lamp issues.

Rae MultiRAE Specifications

General IconGeneral
Operating Temperature-20°C to 50°C (-4°F to 122°F)
Humidity0% to 95% relative humidity (non-condensing)
BatteryRechargeable Li-ion battery
VOC measurement range0-5000 ppm
SensorsPID, LEL, O2, CO, H2S
Battery LifeUp to 18 hours
DisplayBacklit LCD
Data LoggingYes
CommunicationWireless (optional), Bluetooth
AlarmsAudible, visual, and vibrating alarms
PumpInternal pump
CertificationsATEX, IECEx, UL, CSA