83
16 Specifications
Instrument Specifications
Size
7.6" H x 3.8" W x 2.6" D (193 x 96.5 x 66 mm)
Weight
Pumped models: 31 oz. (880 g)
Diffusion models: 26.8 oz. (760 g)
Sensor
Over 25 intelligent interchangeable field-replaceable sensors, including Gamma
radiation, ppb and ppm PIDs, electrochemicals for toxics and oxygen, combustible
LEL and NDIR, and CO
2
NDIR
Battery Options
Rechargeable Li-ion (over 12 hours runtime, pumped/over 18 hours, diffusion;
< 6 hours recharge time)
Extended-duration rechargeable Li-ion (over 18 hours runtime, pumped;
over 27 hours, diffusion)
Alkaline adapter for 4 x AA batteries (approximately 6 hours runtime, pumped/
8 hours, diffusion)
Display
Monochrome graphical LCD display (128 x 160) with backlighting (activated
automatically in dim ambient lighting conditions, when monitor is in alarm, or with
a button press)
Automatic screen flipping.
Display
Readout
Real-time reading of gas concentrations; PID measurement gas and correction
factor; battery status; datalogging on/off; wireless on/off and reception quality.
STEL, TWA, peak, and minimum values
Various instrument status-related information
Keypad
3 operation and programming keys (MODE, Y/+, and N/-)
Sampling
Built-in pump or diffusion. Average flow rate, pumped: 250 cc/min. Auto shutoff in
low-flow conditions
Calibration
Automatic with AutoRAE 2 Test and Calibration Station or manual
Alarms
Wireless remote alarm notification; multi-tone audible (95 dB @ 30 cm),
vibration, visible (flashing bright red LEDs), and on-screen indication of alarm
conditions
Man Down Alarm with pre-alarm and real-time remote wireless notification
Datalogging
Continuous datalogging (six months for 5 sensors at 1-minute intervals, 24/7)
User-configurable datalogging interval (from 1 to 3,600 seconds)
Communication
and
Data Download
Data download and instrument set-up and upgrades on PC via charging and PC
communication cradle, Travel Charger, or AutoRAE 2 Automated Test and
Calibration Station
1
Wireless data and status transmission via built-in RF modem (optional)
Wireless
Network
RAE Systems Dedicated Wireless Network
Wireless
Frequency
ISM license-free bands
EM Immunity
No effect when exposed to 0.43mW/cm
2
RF interference (5-watt transmitter at 12")
Operating
Temperature
-4° to 122° F (-20° to 50° C)