EasyManua.ls Logo

Rae MultiRAE - Span Value Configuration

Rae MultiRAE
93 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MultiRAE User’s Guide
36
8.3.2.10 Change Span Value
You can individually set the span gas concentration for each sensor. This concentration setting will also
be used for a bump test. The units of measure (ppm, %LEL, etc.) are shown on the display.
1. Scroll down the list of sensors using the [N/-] key.
2. Press [Y/+] to select it.
3. Press [N/-] to step through the digits.
4. Press [Y/+] to increase the number from 0 through 9. Once the number 9 is reached, pressing
[Y/+] causes the numbers towrap around” to 0 and count up again.
5. Once you have set the desired value, press [MODE] for “Done.” This registers the new span value.
8.3.2.11 Change Span2 Value
If your MultiRAE is equipped with a high-range, or parts-per billion, PID sensor, you can set the span gas
value for a third calibration point (Span2). The unit of measure is shown on the display.
1. Press [Y/+] to select the highlighted sensor (VOC).
2. Press [N/-] to step through the digits.
3. Press [Y/+] to increase the number from 0 through 9. Once the number 9 is reached, pressing
[Y/+] causes the numbers towrap around” to 0 and count up again.
4. Once you have set the desired value, press [MODE] for “Done.This registers the new Span 2 value.
Note: Three-point calibration is disabled by default, but can be enabled on MultiRAE and MultiRAE Pro
instruments with 10.6eV PID sensors, including high-range ppm and ppb PID sensors. The MultiRAE
Lite PID sensor only supports 2-point calibration.

Table of Contents

Other manuals for Rae MultiRAE

Related product manuals