MultiRAE User’s Guide
80
Search Mode
Alarm Type
Buzzer & LED
Display
Vibration
Reading
Backlight
Priority
Super Alarm
4 beeps/sec
“Super Alarm
screen
400ms
-
On
Highest
Man Down
Alarm
3 beeps/sec
“Man Down Alarm”
screen
400ms
-
On
Man Down
Warning
2 beeps/sec
“Are You OK?
screen
400ms
-
On
Fail
3 beeps/sec
“Lamp” at PID
location
“Off” at LEL
location
400ms
Blinking
reading
On
Pump
3 beeps/sec
Blinking pump
symbol
400ms
Reading
On
Max
3 beeps/sec
“Max” at sensor
location
400ms
Blinking
reading
On
Over Range
3 beeps/sec
“Over” at sensor
location
400ms
Blinking
9999
On
Geiger Counter-style
Alarm
G7 (>High)
7 beeps(30ms)/sec
No change
400ms
Reading
On
G6
6 beeps(40ms)/sec
G5
5 beeps(50ms)/sec
G4
4 beeps(60ms)/sec
G3
3 beeps(70ms)/sec
G2
2 beeps(80ms)/sec
G1 (>Low)
1 beep(90ms)/sec
Negative
1 beep/sec
“Neg” at sensor
location
400ms
0
On
STEL
1 beep/sec
“STEL” at sensor
location
400ms
Reading
On
TWA
1 beep/sec
“TWA” at sensor
location
400ms
Reading
On
Calibration Fail
1 beep/sec
“Cal” at sensor
location
400ms
Reading
On
Bump Fail
1 beep/sec
Bump” at sensor
location
400ms
Reading
On
Datalog Full
1 beep/sec
Blinking datalog
symbol
400ms
Reading
On
Calibration
Required
-
“Full” bottle symbol
-
Reading
-
Bump Required
-
“Empty” bottle
symbol
-
Reading
-
Battery
1 beep/min
Blinking battery
symbol
400ms
Reading
Stays as
is
Nwk Lost
1 beep/min
Blinking RF offline
symbol
400ms
Reading
On
Nwk Joined
1 beep
RF symbol with
RSSI
400ms
Reading
On
Comfort Beep
1 beep/min
No LED flashing
-
-
Reading
-
Lowest

Table of Contents