10
Specifications
These manufacturer’s specifications should be used as guide-
lines when establishing your own performance specification in
accordance with ISO 8655.
Pipet-Lite Specifications
Volume Increment Accuracy Precision
Model µL µL % µL (±) % µL (≤)
2 µL 0.2 0.002 12.0 0.024 6.0 0.012
1.0 2.7 0.027 1.3 0.013
2.0 1.5 0.030 0.7 0.014
10 µL 1.0 0.02 2.5 0.025 1.2 0.012
5.0 1.5 0.075 0.6 0.03
10.0 1.0 0.1 0.4 0.04
20 µL 2 0.02 7.5 0.15 2.0 0.04
10 1.5 0.15 0.5 0.05
20 1.0 0.2 0.3 0.06
100 µL 10 0.2 3.5 0.35 1.0 0.1
50 0.8 0.4 0.24 0.12
100 0.8 0.8 0.15 0.15
200 µL 20 0.2 2.5 0.5 1.0 0.2
100 0.8 0.8 0.25 0.25
200 0.8 1.6 0.15 0.3
1000 µL 100 2 3.0 3.0 0.6 0.6
500 0.8 4.0 0.2 1.0
1000 0.8 8.0 0.15 1.5
2000 µL 200 2 3.0 6.0 0.6 1.2
1000 0.8 8.0 0.2 2.0
2000 0.8 16.0 0.12 2.4
5000 µL 500 5 2.4 12.0 0.6 3.0
2500 0.6 15.0 0.2 5.0
5000 0.6 30.0 0.16 8.0
10 mL 1 mL 20 5.0 50.0 0.6 6.0
5 mL 1.0 50.0 0.2 10.0
10 mL 0.8 80.0 0.16 16.0
20 mL 2 mL 20 5.0 100.0 0.6 12.0
10 mL 1.0 100.0 0.2 20.0
20 mL 0.8 160.0 0.16 32.0
Specifications are subject to change without notice.