V06 - 14 -
Electrical connection
Follow the installation instructions and the infor-
mation on the rating plate when connecting the
unit
The units must be connected to a standard supply
line in conformity with your local regulations.
Comply with all local regulations and standards!
We recommend the use of an earth leakage
circuit breaker (30mA).
On-site installation: provide accessible all-pole
disconnection device with a minimum of a 3 mm
contact gap.
A permanent electrical connection must be
provided for the units.
The thermo cabinet (TC) will be delivered with a
power cord without plug. The connection box is
located on the bottom of the TC. pic 3 A
Connect the power cord in the following way:
PE (yellow/green wire) earth
N (blue wire) neutral
L1 (brown wire) phase
Also there is a connection bolt for the earth con-
nection, located at the rear of the left front leg.
Connect the TC with a suitable cable to the earth
bonding system pic. 3 B
GB
3
A
B
Electrical connection