EasyManua.ls Logo

red lion PAXLA - Module 2: Secondary Function Parameters

red lion PAXLA
12 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7
USER INPUT ASSIGNMENT
Select the value(s) to which the User Input Function is assigned. The User
Input Assignment only applies if a selection of reset, or display hold is selected
in the User Input Function menu.
MODEDISPLAY
No Function

DESCRIPTION
User Input disabled.
Program Mode Lock-out

Zero Input
(Edge triggered)

Zero the Input Display value causing
Display Reading to be Offset.




Resets the assigned value(s) to the current
input value.


See Programming Mode Access chart
(Module 3).
Setpoint 1 and 2 Reset
Setpoint 1 Reset
Setpoint 2 Reset
Reset both setpoint 1 and 2 outputs.
Resets setpoint 1 output.
Resets setpoint 2 output.



USER INPUT FUNCTION




Display Select
(Edge Triggered)
Display Intensity Level
(Edge Triggered)
Display Hold
Reset (Edge triggered)
Advance once for each activation.
Increase intensity one level for each
activation.
Holds the assigned display, but all other
meter functions continue as long as
activated (maintained action).


USER INPUT ACTIVE LEVEL
Select whether the user input is configured as active low or active high.

Entering Code 66 will overwrite all user settings with
the factory settings. The meter will display  and then
return to  . Press the PAR button to exit the
module.
MIN DISPLAY ENABLE







MIN CAPTURE DELAY TIME
When the Input Display is below the present MIN value for the entered delay
time, the meter will capture that display value as the new MIN reading. A delay
time helps to avoid false captures of sudden short spikes.
 to  sec.
Select  to perform any of the Factory Service Operations shown below.
FACTORY SERVICE OPERATIONS

Enables the Minimum Display Capture capability.
The PAXLA uses stored calibration values to provide
accurate measurements. Over time, the electrical
characteristics of the components inside the PAXLA will
slowly change with the result that the stored calibration
values no longer accurately define the input circuit. For
most applications, recalibration every 1 to 2 years should be sufficient.
Calibration of the PAXLA involves a calibration which should only be
performed by individuals experienced in calibrating electronic equipment. Allow
30 minute warm up before performing any calibration related procedure. The
following procedures should be performed at an ambient temperature of 15 to 35
°C (59 to 95 °F).
CAUTION: The accuracy of the calibration equipment will directly affect the
accuracy of the PAXLA.
CALIBRATION


RESTORE FACTORY DEFAULT SETTINGS
Entering Code 50 will display the version (x.x) of the
meter. The display then returns to CodE 00. Press the PAR
button to exit the module.


VIEW MODEL AND VERSION DISPLAY


5.2 MODULE 2 - SECONDARY FUNCTION PARAMETERS ()
PAR
Max Capture
Delay Time
Max Display
Enable
Min Display
Enable
Access Code
For Service
Operations
Min Capture
Delay TIme
Factory
Service
Operations
2-SEC
HI-En HI-t
LO-En LO-t FCS
CodE
Pro
PARAMETER MENU


MAX CAPTURE DELAY TIME
When the Input Display is above the present MAX value for the entered
delay time, the meter will capture that display value as the new MAX reading.
A delay time helps to avoid false captures of sudden short spikes.
 to  sec.


MAX DISPLAY ENABLE

Enables the Maximum Display Capture capability.

Other manuals for red lion PAXLA

Related product manuals