5.3 MODULE 3 - DISPLAY AND FRONT PANEL BUTTON
P
ARAMETERS ()
PAR
Pro
Front Panel
Display
Select Enable
Display
Update Time
Front Panel
Reset Enable
Programming
Security Code
Zero Display
W/Display
Reset
3-dSP
dSP-t SEL rSt ZErO
CodE
Display
Intensity
Level
d-LEV
ScroL
Display
Scroll
Enable
b-LIt
Units Label
Backlight
PARAMETER MENU

DISPLAY UPDATE TIME
This parameter sets the display update time in seconds.

seconds
The  selection allows the SEL button to toggle through the enabled
displays.
FRONT PANEL DISPLAY SELECT ENABLE (SEL)


 
This parameter enables the RST button or user input to zero the input display
value, causing the display reading to be offset.
Note: For this parameter to operate, the RST button or User Input being used
must be set to  and the Input value must be displayed. If these conditions are
not met, the display will not zero.
ZERO DISPLAY WITH DISPLAY RESET


 
The  selection allows the display to automatically scroll through the
enabled displays. The scroll rate is every 4 seconds. This parameter only appears
when the MAX or MIN displays are enabled.
DISPLAY SCROLL ENABLE


 
Enter the desired Display Intensity Level (1-3). The display will actively dim
or brighten as levels are changed.
DISPLAY INTENSITY LEVEL
to

This selection allows the RST button to reset the selected value(s).
FRONT PANEL RESET ENABLE (RST)





 


UNITS LABEL BACKLIGHT*
The Units Label Kit Accessory contains a sheet of custom unit overlays
which can be installed in to the meters bezel display assembly. The backlight
for these custom units is activated by this parameter.
 
Current Calibration
1. Connect the negative lead of a precision DC current source with an accuracy
of 0.01% or better to the COMM terminal. Leave the positive lead of the DC
current source unconnected.
2. With the display at  , press the PAR button. Unit will display 

3. Press the RST button to select the range to be calibrated.
4. Press the
PAR button. Display reads 
5. With the positive lead of the DC current source unconnected, press PAR.
Display reads  for about 8 seconds.
6. When the display reads the selected range, connect the positive lead of the
DC
current source to the current input and apply full-scale input signal for the
range. (Note: For 200 mA range, apply 100 mA as indicated on the display.)
Press PAR. Display reads  for about 8 seconds.
7. Repeat steps 3 through 6 for each input range to be calibrated. When display
reads , press the PAR button to exit calibration.
Voltage Calibration
1. Connect a precision DC voltage source with an accuracy of 0.01% or better
to the volt input and COMM terminals of the PAXLA. Set the output of the
voltage source to zero.
2. With the display at , press the PAR button. Unit will display .
3. Press the RST button to select the range to be calibrated.
4. Press the PAR button. Display reads .
5. With the voltage source set to zero (or a dead short applied to the input), press
PAR. Display reads  for about 8 seconds.
6. When the display reads the selected range, apply full-scale input signal for
the range. (Note: For 200V range, apply 100V as indicated on the display.)
Press PAR. Display reads  for about 8 seconds.
7. Repeat steps 3 through 6 for each input range to be calibrated. When display
reads , press the PAR button to exit calibration