Calibration Procedure



meter should be calibrated by following steps 1 through 7 below.
1. Remove the windshield ball.
2. Power on the meter.
3. Select "50 - 100dB" range.
4. Select "FAST" time weighting.
5. Select "A" weighting.
6. Insert Microphone into the sound calibrator's microphone adapter.
7. Power on the sound calibrator (R8090).
If the measured value is not within "
calibration potentiometer until the display matches the output of the
calibrator (94.0dB).
Data Stream Output



Terminal Settings:




41170100000715 =
+71.5 dB
Reading
Decimal Place 0-3
Positive (+) = 0, Negative (-) = 1
dB = 17
1
4
continued...