6
 
Data Output: Yes (RS-232)
AC Output Signal: 0.5Vrms (full scale of selected range)
AC Output Impedance: 600 ohms
AC Output Terminal: 3.5mm Stereo Jack
Battery Life: Sampling Time Dependent
Product Certications: CE, IEC 61672-1-2013, Class 2
Operating Temperature: 32 to 122°F (0 to 50°C)
Storage Temperature: 14 to 140°F (-10 to 60°C)
 
 
Weight: 0.7lbs (320g)
Comparison of Frequency Weightings and Tolerance Limits
Frequency Hz
A Weighting
Characteristics
C Weighting
Characteristics
Tolerance
(IEC 61672 class 2)
31.5 -39.4dB -3dB 
63 -26.2dB -0.8dB 
125 -16.1dB -0.2dB 
250 -8.6dB 0dB 
500 -3.2dB 0dB 
1K 0dB 0dB 
2K +1.2dB -0.2dB 
4K +1dB -0.8dB 
8K -1.1dB -3dB 
Time Weighting (Fast & Slow) Characteristics
Time Weighting
Characteristics
Max Response Reference
Continuous Signal
Tolerance
(IEC 61672 class 2)
F (Fast) -1.0dB 
S (Slow) -4.1dB 