EasyManua.ls Logo

ReliOn Confirm - Perform Control Solution Test

ReliOn Confirm
74 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
24
Perform a Control Solution Test:
•BeforetestingwiththeReliOn®Conrmbloodglucosemeterfor
the first time.
•Wheneveryoususpectthemeterorteststripsmaynotbe
functioning properly.
•Iftestresultsappeartobeabnormallyhighorloworarenot
consistent with clinical symptoms.
•Theteststripbottlehasbeenleftopenorhasbeenexposedto
temperatures below 34°F (1°C) or above 86°F (30°C), or humidity
levels above 80%.
•Tocheckyourtechnique.
•WhentheReliOn®Conrmbloodglucosemeterhasbeen
dropped or stored below 32°F (0°C) or above 122°F (50°C).
Important:
•UseonlywithReliOn®Conrmbloodglucosemeterand
ReliOn® Confirm/micro blood glucose test strips.

Related product manuals