•TheReliOn®ConrmBloodGlucoseMonitoringSystemisnot
designed to be a substitute for pathology laboratory equipment and
should not be used for the diagnosis of diabetes.
•Useonlyfreshcapillaryblood.Donotuseserumorplasmaorvenous
whole blood.
•DonotusetheReliOn®ConrmBloodGlucoseMonitoringSystem
to test neonates. It has not been validated for neonatal use.
•Alwaysinserttheteststripintothemeterrst,thenprickthenger.
Not doing so may result in an“E-6” appearing on the display or a test
result that is higher than the actual value.
•Donotuseataltitudeshigherthan10,000ft(3,048m).