EasyManua.ls Logo

REPLIGEN OPUS 14 - Figure 4. Sanisure Clamp Removal; Figure 5. Stainless Steel Valve Connection to OPUS Column

REPLIGEN OPUS 14
20 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
UG-OPU 5-80R
11
OPUS® 5-80R Pre-packed Chromatography Columns
b. Reference Table 1 and Table 7 in this User Guide for handling suggestions and features.
c. Do not pull or strain the white inlet and outlet ports protruding from the top of the
OPUS® column.
4. If you are not ready to use the column, refer to the storage recommendations of the resin
supplier for the pre-packed resin.
a. OPUS® 8-80R Columns, refer to column label for recommended storage temperature.
b. OPUS® 5 Column label does not report recommended storage temperature.
5. When you are ready to use your OPUS®
Column, remove the clear plastic bag containing the
OPUS® 5 30 Column. There is no bag for OPUS®45R, 60R and 80R Columns, so they should
be wiped down prior to use with Ethanol or Isopropyl Alcohol solution if necessary.
a. If storing the column in 2-8
o
C, allow the column to equilibrate to room temperature
overnight prior to use of the column.
b. Using a wire cutter or sharp scissors, remove the white cable-tie on the SaniSure
®
clamp
sealing the inlet port. See Figure 4
Figure 4. SaniSurClamp Removal
Connecting and Operating Your OPUS® Column
Notes:
Use of stainless steel valves connected directly to the inlet and/or the outlet ports is not
recommended. The additional weight to the top of the column will increase the risk of damage to
the column hardware.
If valves must be used, first connect the inlet and outlet ports to tubing. Then the other end
of that tubing may be connected to valves as seen in Figure 5
Tubing length of <50 cm is suggested to minimize hold up volume of the system
Figure 5. Stainless Steel Valve Connection to OPUS® Column

Table of Contents