OPUS® 5-80R Pre-packed Chromatography Columns
Reduce the flow rate of solution through the column to abide by pressure limit of the
packed resin bed
Confirm that high viscosity solutions are not being used during pressure evaluation
Flow of alcohols through the column is known to increase column pressure
Allow all buffers and the column to equilibrate to ambient temperatures
5.3 Pressure increase during run
Causes
Product or precipitates clogging of the polypropylene mesh
Operation under higher flow rate than recommended for the packed resin
Residue build up at the top of the column
Use of high viscosity solutions or high product load concentrations
Fouled chromatography resin
Temperature shifts between buffers used in the column
Potential Fixes
Clean the column with the appropriate cleaning method for the residue that clogged the
mesh and/or resin. Running in reverse flow, or up-flow mode is recommended
Flow >5CV of equilibration buffer through the column in reverse flow. Recheck pressure and
column performance (Efficiency, Asymmetry, Pressure vs. Flow) under normal operating
conditions for comparison to results for on CoA
Check valve position
Allow all buffers and the column to equilibrate to ambient temperatures
5.4 Pressure drop during run
Causes
Line or fitting leaks
Temperature shifts during buffer transitions
Viscosity shifts during buffer transitions
Potential Fixes
Check lines and connections
Allow all buffers and the column to equilibrate to ambient temperatures