• Reduce the flow rate of solution through the column to abide by pressure limit of the
packed resin bed
• Confirm that high viscosity solutions are not being used during pressure evaluation
• Flow of alcohols through the column is known to increase column pressure
• Allow all buffers and the column to equilibrate to ambient temperatures
5.3 Pressure increase during run
Causes
• Product or precipitates clogging of the polypropylene mesh
• Operation under higher flow rate than recommended for the packed resin
• Residue build up at the top of the column
• Use of high viscosity solutions or high product load concentrations
• Fouled chromatography resin
• Temperature shifts between buffers used in the column
Potential Fixes
• Clean the column with the appropriate cleaning method for the residue that clogged the
mesh and/or resin. Running in reverse flow, or up-flow mode is recommended
• Flow >5CV of equilibration buffer through the column in reverse flow. Recheck pressure and
column performance (Efficiency, Asymmetry, Pressure vs. Flow) under normal operating
conditions for comparison to results for on CoA
• Check valve position
• Allow all buffers and the column to equilibrate to ambient temperatures
5.4 Pressure drop during run
Causes
• Line or fitting leaks
• Temperature shifts during buffer transitions
• Viscosity shifts during buffer transitions
Potential Fixes
• Check lines and connections
• Allow all buffers and the column to equilibrate to ambient temperatures