Supplementary features
98
(A)CV Breath pattern
Parameter Setting
Vt (mL)
Adult: 100 to 2,500 [500]
Paed: 50 to 300 [100]*
Ti (sec)
When Volume Breath option is set to Ti:
Adult: 0.3 to 3.0 [1.0]
Paed: 0.3 to 3.0 [0.6]
PIF (L/min)
When Volume Breath option is set to PIF:
Adult: 10 to 120 [50]
Paed: 5 to 60 [10]
Resp. rate (per min)
Adult: 4 to 50 [15]
Paed: 12 to 80 [15]
*The International ventilator standard indicates that Paediatric patient type is intended to be used for a patient receiving less
than 300 mL, however Astral permits adjustment of ‘Vt’ setting parameter up to 500 mL for cases where ‘Vt’ is set such that it
compensates for leak in the breathing circuit.
WARNING
ResMed does not recommend 500 mL as the upper limit for the pediatric tidal volume use;
however, clinicians may choose this upper limit based on their clinical determination.

Table of Contents

Other manuals for ResMed Astral Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ResMed Astral Series and is the answer not in the manual?

ResMed Astral Series Specifications

General IconGeneral
TypeVentilator
Displaycolor LCD
ConnectivityUSB
EPAP0 to 25 cm H2O
Rise TimeAdjustable
PEEP0 to 25 cm H2O
FiO2 MonitoringYes, with optional oxygen module
Battery BackupInternal and external battery options
Oxygen CapabilityYes, with optional oxygen module
AlarmsLow battery
Pressure Support0 to 30 cm H2O

Related product manuals