Technical specifications
138
Technical specifications
Operating pressure range
Single limb with valve or double limb with valve: 3 to 50 hPa
Single limb with intentional leak: 2 to 50 hPa
CPAP: 3 to 20 hPa
Maximum working pressure limit: 10 to 80* hPa
*Not applicable on all device variants
Forced cycling occurs if the Pressure alarm limit is exceeded.
Operating tidal volume range (volume
control modes)
Adult patient type: 100 to 2500 mL
Paediatric patient type: 50 to 300 mL*
Circuit resistance and compliance range
for stated accuracy of monitoring and
control**
Paediatric patient setting:
Circuit resistance range (circuit with intentional leak): 0 to 8 hPa at
60 L/min
Circuit resistance range (circuit with valve): 0 to 20 hPa at 60 L/min
Circuit resistance range (mouthpiece circuit): 0 to 5 hPa at 60 L/min
Circuit compliance range: 0 to 4 mL / hPa
Adult patient setting:
Circuit resistance range (circuit with intentional leak): 0 to 20 hPa at
120 L/min
Circuit resistance range (circuit with valve): 0 to 35 hPa at 120 L/min
Circuit resistance range (mouthpiece circuit): 0 to 15 hPa at 120 L/min
Circuit compliance range: 0 to 4 mL / hPa
Breathing resistance under single
fault***
Paediatric circuit
Inspiration: 2.2 hPa (at 15 L/min), 5.3 hPa (at 30 L/min)
Expiration: 2.4 hPa (at 15 L/min), 5.0 hPa (at 30 L/min)
Adult circuit
Inspiration: 5.7 hPa (at 30 L/min), 8.3 hPa (at 60 L/min)
Expiration: 4.2 hPa (at 30 L/min), 6.2 hPa (at 60 L/min)
Maximum flow
220 L/min
Inspiratory trigger (nominal)
characteristics
Inspiratory trigger occurs when patient flow exceeds trigger setting.
Double limb with valve (flow trigger): 0.5 to 15.0 L/min
Single limb with valve or double limb with valve: 1.6 to 10.0 L/min (in five
steps)****
Single limb with intentional leak: 2.5 to 15.0 L/min (in five steps)
Mouthpiece (tube only) circuit: 2.0 to 4.0 L/min (in four steps)
Expiratory cycle (nominal)
characteristics
Cycle occurs when inspiratory flow declines to the set percentage of peak
inspiratory flow.
5 to 90%
Sound pressure level
35 ± 3 dBA as measured according to ISO 80601-2-72:2015
Sound power level
43 ± 3 dBA as measured according to ISO 80601-2-72:2015
Alarm volume range
5685 dBA (in five steps) as measured according to IEC 60601-1-8:2012
Data storage
7 days of high-resolution airway pressure, respiratory flow and delivered
volume (sampled at 25 Hz).
7 days of breath-related therapy data (sampled at 1 Hz).
365 days of statistical data per program.

Table of Contents

Other manuals for ResMed Astral Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ResMed Astral Series and is the answer not in the manual?

ResMed Astral Series Specifications

General IconGeneral
TypeVentilator
Displaycolor LCD
ConnectivityUSB
EPAP0 to 25 cm H2O
Rise TimeAdjustable
PEEP0 to 25 cm H2O
FiO2 MonitoringYes, with optional oxygen module
Battery BackupInternal and external battery options
Oxygen CapabilityYes, with optional oxygen module
AlarmsLow battery
Pressure Support0 to 30 cm H2O