Monitoring
144
Parameter
Updates
Units
Range
Resolution
Accuracy
Measured Unintentional
Leak
(for double limb circuit
with expiratory valve)
End of breath
%
0 to 100
1
N/A
* When Resp. rate ≥ 8/min, and Pressure ≤ 30 hPa for circuits with intentional leak, and with non-compliant interface.
Pressure parameters
Parameter
Updates
Units
Range
Resolution
Accuracy
Pressure
Continuous
cmH
2
O or hPa
or mbar
0 to 99
0.1
±(0.5 hPa + 4% of
actual pressure)
PIP
End of
inhalation
cmH
2
O or hPa
or mbar
0 to 99
0.1
±15%
PEEP
End of
exhalation
cmH
2
O or hPa
or mbar
0 to 30
0.1
±(0.5 hPa + 4% of
actual pressure)
Avg. P
End of breath
cmH
2
O or hPa
or mbar
0 to 99
0.1
±(0.5 hPa + 4% of
actual pressure)
Other parameters
Parameter
Updates
Units
Range
Resolution
Accuracy
RSBI
End of breath
1/min-L
0 to 999
1
N/A
% Spont. Trig.
Start of
inhalation
%
0 to 100
1
N/A
% Spont. Cyc.
End of
inhalation
%
0 to 100
1
N/A
Measured Battery Time
Once per
minute
HH:MM
N/A
1 min
N/A
Measured Battery Charge
Level
Once per
minute
%
0 to 100
1
N/A
FiO
2
*
End of
inhalation
%
18 to 100
1
±(2.5% + 2.5% of
actual oxygen
concentration)**
SpO
2
Once per
second
% SpO
2
0 to 100
1
Refer to Nonin
Xpod 3012
technical
specifications at
www.nonin.com
Pulse rate
Once per
second
1/min
18 to 321
1
Refer to Nonin
Xpod 3012
technical
specifications at
www.nonin.com.
* FiO
2
monitoring automatically compensates for atmospheric pressure variation.
** The measured FiO
2
sampled at 25 Hz has a response time of <12s to 90% of the final value. The long term output drift is <1% volume oxygen
per month.

Table of Contents

Other manuals for ResMed Astral Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ResMed Astral Series and is the answer not in the manual?

ResMed Astral Series Specifications

General IconGeneral
TypeVentilator
Displaycolor LCD
ConnectivityUSB
EPAP0 to 25 cm H2O
Rise TimeAdjustable
PEEP0 to 25 cm H2O
FiO2 MonitoringYes, with optional oxygen module
Battery BackupInternal and external battery options
Oxygen CapabilityYes, with optional oxygen module
AlarmsLow battery
Pressure Support0 to 30 cm H2O