21.11.2001 Retsch GmbH 8 Doc. No E98.014-016.9999 A
Requirements for the installation site
Ambient temperature
5°C to 40°C
When the ambient temperature exceeds or falls below that
specified, the electronic and mechanical components may
be damaged, and performance data changed to an
unknown extent.
Humidity:
Maximum relative humidity 80% at temperatures up to 31°C;
linear decline down to 50% relative humidity at 40°C.
At higher humidity, the electronic and mechanical
components may be damaged, and performance data
changed to an unknown extent.
Installation site – altitude:
max. 2000 m above mean sea level
Installation / Shipping nuts
SM
Set up the AS200 only on a stable laboratory bench to avoid un-
pleasant transmission of vibrations.
Loosen the hexagonal nuts SM using an open-ended spanner until
the sieve plate moves freely
Unscrew the wing screws A (shipping nuts) with hexagonal nuts.
Mount the appropriate sieve clamping device
(See the chapter
on operation for instructions)
Keep the wing screws A with hexagonal nuts (shipping nuts) in a
safe place for later transportation.
Reusing the shipping nuts
Screw the hexagonal nuts fully onto the wing screws
Screw the wing screws through the sieve plate into the housing
Twist the hexagonal nuts downwards and tighten until the sieve
plate can no longer move
Operation with these shipping nuts in place, or shipping
without the these nuts installed, could cause damage to
mechanical components.