RIGOL     DG1000 Command System                                
          
 Programming Guide for DG1000 
Detailed information of each command: 
 
1.
APPLy:SINusoid 
Syntax  APPLy:SINusoid [<frequency>[,<amplitude>[,<offset>]]] 
Function  Generate a sine wave with specific frequency, amplitude and DC 
offset via CH1.  
Explanations    If the parameters you set are less than three, the sequence 
would be: <frequency>, <amplitude>, <offset>.  
  The default units of <frequency>, <amplitude> and <offset> 
are Hz, Vpp and V
 respectively. 
Example
APPL:SIN 1000,5.0,-1.5 
 
2.
APPLy:SQUare 
Syntax  APPLy:SQUare [<frequency>[,<amplitude>[,<offset>]]] 
Function  Generate a square wave with specific 
frequency, amplitude and DC 
offset via CH1. This command would overwrite the current duty cycle 
setting and select 50% automatically.  
Explanations    If the parameters you set are less than three, the sequence 
would be: <frequency>, <amplitude>, <offset>.  
  The default units of <frequency>, <amplitude> and <offset> 
are Hz, Vpp and V
DC
 respectively. 
Example
APPL:SQU 1000,5.0,-1.5 
 
Syntax
APPLy:RAMP [<frequency>[,<amplitude>[,<offset>]]]
Function  Generate a ramp wave with specific 
frequency, amplitude and DC 
offset via CH1. This command would overwrite the current symmetry 
setting and select 50% automatically. 
Explanations    If the parameters you set are less than three, the sequence 
would be: <frequency>, <amplitude>, <offset>.  
  The default units of <frequency>, <amplitude> and <offset> 
are Hz, Vpp and V
 respectively. 
Example  APPL:RAMP 1000,5.0,-1.5 
 
4.
APPLy:PULSe 
Syntax  APPLy:PULSe [<frequency>[,<amplitude>[,<offset>]]] 
Function
Generate a pulse wave with specific 
frequency, amplitude and DC