RIGOL Chapter 2 Performance Verification Test
DG1000Z Performance Verification Guide
2-8
respectively and judge whether the readings are within the specification
(“Offsetin Table 2-2) range.
Table 2-2 Offset limits of DC offset accuracy test
Offset Setting
Value
Amplitude
Setting Value
Allowed Error
[1]
Offset
[2]
-2.5V
DC
5Vpp
±0.005V
DC
-2.505V
DC
to -2.495V
DC
-1V
DC
±0.020V
DC
-1.02V
DC
to -0.98V
DC
-500mV
DC
±0.025V
DC
-0.525V
DC
to -0.475V
DC
0V
DC
±0.030V
DC
-0.030V
DC
to 0.030V
DC
500mV
DC
±0.035V
DC
0.465V
DC
to 0.535V
DC
1V
DC
±0.040V
DC
0.96V
DC
to 1.04V
DC
2.5V
DC
±0.055V
DC
2.445V
DC
to 2.555V
DC
Note
[1]
: Allowed Error” is calculated from the specification “± (1% of setting value + 5 mV + 0.5% of
amplitude)”.
Note
[2]
: Offset = offset setting value ± allowed error.
7. Repeat steps 1 to 6 to test the DC offset accuracy of CH2 and record the test
results.
Test Record Form:
Offset
Setting
Value
Setting
Measurement
Value
Specification Pass/Fail
-2.5V
DC
Frequency:
1kHz
Amplitude:
5Vpp
Impedance:
50Ω
-2.505V
DC
to -2.495V
DC
-1V
DC
-1.02V
DC
to -0.98V
DC
-500mV
DC
-0.525V
DC
to -0.475V
DC
0V
DC
-0.030V
DC
to 0.030V
DC
500mV
DC
0.465V
DC
to 0.535V
DC
1V
DC
0.96V
DC
to 1.04V
DC
2.5V
DC
2.445V
DC
to 2.555V
DC

Other manuals for Rigol DG1032Z