Chapter 5 Specifications RIGOL
DG1000Z Users Guide 5-7
Environment
Temperature Range
Operating: 0 to 50
Non-Operating: -40 to 70
Cooling Method Cooling by fans compulsively
Humidity Range
Less than 30: ≤95% Relative Humidity (RH)
30 to 40: ≤75% Relative Humidity (RH)
40 to 50: ≤45 Relative Humidity (RH)
Altitude
Operating: Less than 3000 meters
Non-Operating: Less than 15,000 meters
Mechanical
Dimensions
(W×H×D)
261.5mm×112mm×318.4mm
Weight
without package: 3.2kg
with package: 4.5kg
Interfaces USB Host, USB Device, LAN
IP Protection
IP2X
Calibration
Interval
Recommend calibration interval is one year
Authentication Information
EMC
In line with
EN61326-1:2006
IEC 61000-3-2:2000
±4.0kV (Contact Discharge)
±4.0kV (Air Discharge)
IEC 61000-4-3:2002
3V/m (80MHz to 1GHz)
3V/m (1.4GHz to 2GHz)
1V/m (2.0GHz to 2.7GHz)
IEC 61000-4-4:2004 1kV power lines
IEC 61000-4-5:2001
0.5kV (Phase to Neutral)
0.5kV (Phase to PE)
1kV (Neutral to PE)
IEC 61000-4-6:2003
3V, 0.15-80MHz
EC 61000-4-11:2004
Voltage dip:
0%UT during half cycle
0%UT during 1 cycle
70%UT during 25 cycle
Short interruption:
0%UT during 1 cycle
Electrical Safety
In line with
USA: UL 61010-1:2012,
Canada: CAN/CSA-C22.2 No. 61010- 1-2012
EN 61010-1:2010