Chapter 2 Command System  RIGOL 
DG800 Programming Guide  2-9 
:COUPling[<n>]:AMPL:DEViation 
 
Syntax 
:COUPling[<n>]:AMPL:DEViation <deviation> 
:COUPling[<n>]:AMPL:DEViation? 
 
Description   
Sets the amplitude deviation in the amplitude coupling of the specified channel. 
Queries the amplitude deviation in the amplitude coupling of the specified channel. 
 
Parameter 
-19.998 Vpp to 19.998 Vpp
 
 
Remarks 
  When [<n>] is omitted, the commands set and query the related parameters of CH1 by default. 
  Select the desired amplitude coupling mode (
:COUPling[<n>]:AMPL:MODE) and set the corresponding 
amplitude deviation or amplitude ratio (
:COUPling[<n>]:AMPL:RATio) before enabling the amplitude 
coupling function (
:COUPling[<n>]:AMPL[:STATe]). You cannot set the amplitude coupling mode and 
amplitude deviation/ratio after the amplitude coupling function is enabled. 
  When the amplitude coupling function is disabled, if the current amplitude coupling mode is amplitude 
deviation, sending this command can set the amplitude deviation; if the current amplitude coupling 
mode is amplitude ratio, sending this command can set the amplitude coupling mode to amplitude 
deviation and set the amplitude deviation. 
 
Return Format 
The query returns the amplitude deviation in scientific notation. The returned value contains 7 effective 
digits, for example, 1.000000E+00 (the amplitude deviation is 1 Vpp). 
 
Example   
:COUP1:AMPL:DEV 1    /*Sets the amplitude deviation in the amplitude coupling of CH1 to 1 Vpp.*/ 
:COUP1:AMPL:DEV?     /*Queries the amplitude deviation in the amplitude coupling of CH1 and the 
query returns 1.000000E+00.*/ 
 
 
:COUPling[<n>]:AMPL:MODE 
 
Syntax 
:COUPling[<n>]:AMPL:MODE {OFFSet|RATio} 
:COUPling[<n>]:AMPL:MODE? 
 
Description 
Sets the amplitude coupling mode of the specified channel to OFFSet (amplitude deviation) or RATio 
(amplitude ratio). 
Queries the selected amplitude coupling mode of the specified channel. 
 
Parameter 
 
Remarks   
  When [<n>] is omitted, the commands set and query the related parameters of CH1 by default. 
  Amplitude deviation mode: the amplitudes of CH1 and CH2 have a certain deviation relation. The 
parameter relations are A
CH2
=A
CH1
+A
Dev
 (the reference source is CH1); A
CH1
=A
CH2
-A
Dev
 (the reference 
source is CH2). Wherein, A
CH1
 is the amplitude of CH1, A
CH2
 is the amplitude of CH2 and A
Dev
 is the set