RIGOL
© 2007 RIGOL Technologies, Inc.
User‟s Guide for DM3000 Series
5-9
Inspection Function on Rear Panel (for DM3064 and DM3054)
CAUTION:
The voltage limit that the LO terminal relatives to the chassis ground is 150Vpeak (Max).
Channels: 12 differential voltage channels, 4 differential current channels.
Measurement Functions: 2WR, Capacitance, DCV, DCI, ACV, ACI, Diodes , frequency and Period.
Work characteristic: Thermo emf <6uV. Maximum scanning rate is 2 channels per second.
Input characteristic: Diff input voltage 150Vpeak(Max), isolated voltage between channels 150Vpeak(Max),
current input terminal 1Apeak(Max), channel isolation >60dB(@10kHz), all terminal to the
chassis ground voltage 150Vpeak(Max).
Current channel protection: Inner current channel 2A self- recover fuse
Voltage channel protection: 250V over voltage protection.
Real-time Clock
Precision: 1min/month (Environment Temperature >0)
Clock battery Life: 2 years
Math Functions
Null, Min/Max/Average, dBm, dB, Limit Test (with TTL output)
Other Functions
Reading Hold, Ratio Measurement
High-speed Sampling
50kSa/s (In DataLog function)
Reading Resolution
2,400,000 Count, >6 1/2 digits
USB I/O Interface
USB Host, USB Device interface. It can support U-disk and USB printer.
Other I/O Interface
RS-232, GPIB (Selectable) support for SCPI command, LAN (Selectable)