1-22
To Check Diodes
The following shows the system connections and selection of measurement functions.
The practice provides a guide to get familiar with the Check Diodes technique.
Figure 1- 29 Check Diodes Interface
Table1- 9 Check Diodes Characteristics
Tests Current
1mA
Max Resolution
Range fixed at 2VDC
Open-circuit Voltage
<7V
Input Protection
1000V (HI Terminal)
Configurable
Parameters
0.1V≤V
measured
≤2V
Steps:
1. Connect test leads as shown in Figure 1- 30, red test lead to the HI Terminal,
black test lead to the LO Terminal.
2. Press to select the Check Diodes.