EasyManua.ls Logo

Rigol DM3061

Rigol DM3061
158 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RIGOL
© 2007 RIGOL Technologies, Inc.
User‟s Guide for DM3000 Series
5-12
[1] Specifications are for 60 minute warmup, selecting 5 3/4 reading resolution and calibration temperature
18 28 .
[2] 20% over range on all ranges, except DCV 1000V, ACV 750V, DCI and ACI 10A range.
[3] Specifications are for 4wire resistance function, or 2wire resistance using Math Null. Without Math Null,
add 0.2 Ω additional errors in 2wire resistance function.
[4] For each additional volt over ± 500 VDC add 0.02 mV of error.
[5] For current terminal, > 7A DC or ACRMS for 30 seconds ON and 30 seconds OFF.
[6] Accuracy specifications are for the voltage measured at the input terminals only. 1 mA test current is
typical.
Variation in the current source will create some variation in the voltage drop across a diode junction.
Settling Considerations
Reading settling times are affected by source impedance, cable dielectric characteristics, and input signal
changes. Typically, settling time <1.5s when source impedance less than 1kΩ.

Table of Contents

Other manuals for Rigol DM3061

Related product manuals