User‟s Guide for DM3000 Series
2-4
AC Filter
There are three settings at the AC Filter menu. Choose the appropriate setting for
more accurate measurements. This function applies AC Voltage and AC Current
measurement only.
Press Filter, enter the menu shown below:
Figure 2- 2
Table 2- 2 AC Filter Menu Description
Function
Menu
Description
Slow
Set up the filter with low speed.
Mid
Set up the filter with to middle speed.
Fast
Set up the filter with high speed.
Save all changes, back to a higher level menu.
Table 2- 3 AC Filter Parameters Characteristics
AC Filter Options
Input Frequency
Setting Timer
Slow
3Hz300kHz
1.2 reading/s
Mid
20Hz300kHz
0.5 reading/s
Fast
200Hz300kHz
0.3 reading/s
AC Filter
The AC Filter Parameters are saved in the volatile memory, the data will lose when
the power is off.
The default value of AC Filter Parameters is “Mid” (middle).