Chapter 1 Calibration Notices RIGOL
DP700 Calibration Guide 1-3
Calibration Precautions
To achieve an optimal performance, you should follow the advices below to perform
all the calibration procedures.
Ensure that the ambient temperature stays between 18°C and 28°C. Ideally, the
calibration should be performed at 23°C±1°C.
Ensure that the ambient relative humidity is less than 80%.
Ensure that the instrument has been warmed up for at least 30 minutes before
the calibration.
Keep the cables of the test probe as short as possible.