EasyManua.ls Logo

Rigol DP700 Series

Rigol DP700 Series
21 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 1 Calibration Notices RIGOL
DP700 Calibration Guide 1-3
Calibration Precautions
To achieve an optimal performance, you should follow the advices below to perform
all the calibration procedures.
Ensure that the ambient temperature stays between 18°C and 28°C. Ideally, the
calibration should be performed at 23°C±1°C.
Ensure that the ambient relative humidity is less than 80%.
Ensure that the instrument has been warmed up for at least 30 minutes before
the calibration.
Keep the cables of the test probe as short as possible.

Related product manuals