RIGOL Chapter 2 Command System
2-30 DP800 Programming Guide
:MONItor:VOLTage[:VALue]
Syntax :MONItor:VOLTage[:VALue] {<value>|MINimum|MAXimum}
:MONItor:VOLTage[:VALue]? [MINimum|MAXimum]
Description Set the voltage of the monitor condition.
Query the voltage of the monitor condition.
Parameter
Name Type Range Default
<value> Real
0 to the maximum voltage
of the current channel
0.5*rated voltage of the
current channel
Return Format The query returns the voltage of the current monitor condition, for example, 5.000.
Example :MONI:VOLT 5 /*Set the voltage of the monitor condition to 5V*/
:MONI:VOLT? /*Query the voltage of the current monitor condition and the
query returns 5.000*/
Related
Commands
:MONItor:CURRent:CONDition
:MONItor:CURRent[:VALue]
:MONItor:POWER:CONDition
:MONItor:POWER[:VALue]
:MONItor:VOLTage:CONDition