RIGOL Chapter 2 Constant Voltage Tests
DP800 Performance Verification Manual
2-6
Table 2-2 DP832A CV Load Regulation Rate Test Record Form
Channel CH1 (30V/3A) CH2 (30V/3A) CH3 (5V/3A)
Specification <5mV <5mV <2.5mV
DP832A
Setting
Voltage 30V 30V 5V
Current 3.2A 3.2A 3.2A
Measurement
Result
U
0
U
1
|U
1
-U
0
|
Table 2-3 DP832 CV Load Regulation Rate Test Record Form
Channel CH1 (30V/3A) CH2 (30V/3A) CH3 (5V/3A)
Specification <5mV <5mV <2.5mV
DP832
Setting
Voltage 30V 30V 5V
Current 3.2A 3.2A 3.2A
Measurement
Result
U
0
U
1
|U
1
-U
0
|
Table 2-4 DP821A CV Load Regulation Rate Test Record Form
Channel CH1 (60V/1A) CH2 (8V/10A)
Specification <8mV <2.8mV
DP821A
Setting
Voltage 60V 8V
Current 1.05A 10.5A
Measurement
Result
U
0
U
1
|U
1
-U
0
|
Table 2-5 DP811A CV Load Regulation Rate Test Record Form
Range Range1 (20V/10A) Range2 (40V/5A)
Specification <4mV <6mV
DP811A
Setting
Voltage 20V 40V
Current 10.5A 5.3A
Measurement
Result
U
0
U
1
|U
1
-U
0
|