Command Systems RIGOL
Programming Guide for DS1000B Series
2-19
The command clears the out of date waveforms on the screen during waveform
persist.
8. :DISPlay:BRIGhtness
Command Format:
:DISPlay:BRIGhtness <count>
:DISPlay:BRIGhtness?
Function:
The command sets the brightness of grid. The <count> range is 0~100, and the
bigger the count is, the brighter the grid becomes.
Returned Format:
The query returns 0 or 1, 2 ……100.
Example:
:DISP:BRIG 10 Set the grid brightness as 10.
:DISP:BRIG? Return 10.
9. :DISPlay:INTensity
Command Format:
:DISPlay:INTensity <count>
:DISPlay:INTensity?
Function:
The command sets the brightness of the waveform. The <count> range is
0~100, and the bigger the count is, the brighter the waveform becomes.
Returned Format:
The query returns 0 or 1, 2 ……100.
Example:
:DISP:INT 12 Set the waveform brightness as 12.
:DISP:IN? Return 12.