EasyManuals Logo

Rigol DS1000B Series Programming Guide

Rigol DS1000B Series
152 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #31 background imageLoading...
Page #31 background image
Command Systems RIGOL
Programming Guide for DS1000B Series
2-19
The command clears the out of date waveforms on the screen during waveform
persist.
8. :DISPlay:BRIGhtness
Command Format:
:DISPlay:BRIGhtness <count>
:DISPlay:BRIGhtness?
Function:
The command sets the brightness of grid. The <count> range is 0~100, and the
bigger the count is, the brighter the grid becomes.
Returned Format:
The query returns 0 or 1, 2 ……100.
Example:
:DISP:BRIG 10 Set the grid brightness as 10.
:DISP:BRIG? Return 10.
9. :DISPlay:INTensity
Command Format:
:DISPlay:INTensity <count>
:DISPlay:INTensity?
Function:
The command sets the brightness of the waveform. The <count> range is
0~100, and the bigger the count is, the brighter the waveform becomes.
Returned Format:
The query returns 0 or 1, 2 ……100.
Example:
:DISP:INT 12 Set the waveform brightness as 12.
:DISP:IN? Return 12.

Other manuals for Rigol DS1000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1000B Series and is the answer not in the manual?

Rigol DS1000B Series Specifications

General IconGeneral
BrandRigol
ModelDS1000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals