Command Systems RIGOL
Programming Guide for DS1000B Series
2-39
ALTernation Trigger
1. :TRIGger:ALTernation:SOURce
Command Format:
:TRIGger:ALTernation:SOURce <source>
:TRIGger:ALTernation:SOURce?
Function:
The command selects the alternation trigger channel. The <source> may be
CH1CH2, CH1CH3, CH1CH4, CH2CH3, CH2CH4 or CH3CH4.
Returned Format:
The query returns CH1CH2 or CH1CH3, CH1CH4, CH2CH3, CH2CH4, CH3CH4.
Example:
:TRIG:ALT:SOUR CH1CH2 Set the alternation channel as CH1CH2.
:TRIG:ALT:SOUR? Return CH1CH2.
2. :TRIGger:ALTernation:CURRentSOURce
Command Format:
:TRIGger:ALTernation:CURRentSOURce <source>
:TRIGger:ALTernation:CURRentSOURce?
Function:
The command sets the current channel. The <source> may be SOURceA or
SOURceB, and the source A and B are varying with the current alternation
channel.
Returned Format:
The query returns SOURceA or SOURceB.
Example:
:TRIG:ALT:SOUR CH1CH2 Set the alternation channel as CH1CH2.
:TRIG:ALT:CURRSOUR SOURB Set the current channel as source B.
:TRIG:ALT:CURRSOUR? Return SOURceB.

Other manuals for Rigol DS1000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1000B Series and is the answer not in the manual?

Rigol DS1000B Series Specifications

General IconGeneral
BrandRigol
ModelDS1000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals