:TIMebase[:MAIN]:SCALe <scale_val>
:TIMebase[:MAIN]:SCALe?
Function:
The command sets the timebase scale of main mode, and the unit is s/div.
In NORMAL mode, different types of instruments have different sweep
ranges:
DS1204B , <scale_val> range: 1ns/div~50s/div.
DS1104B, <scale_val>range: 2ns/div~50s/div.
DS1074B, <scale_val>range: 5ns/div~50s/div.
In SCAN mode, <scale_val>range: 50ms ~ 50s.
Returned Format:
The query returns the value of timebase scale, and the unit is s.
Example:
:TIM:MODE MAIN Set the scan mode as main.
:TIM:SCAL 2 Set the timebase scale as 2s.
:TIM:SCAL? Return 2.000e000.
5. :TIMebase:DELayed:SCALe
Command Format:
:TIMebase:DELayed:SCALe <scale_val>
:TIMebase:DELayed:SCALe?
Function:
The command sets the timebase scale of delayed scan, and the unit is s/div.
When the “Delayed” is “ON”, for view waveform details, the waveform may be
amplified under the width of window vary with the delayed timebase scale.
In NORMAL mode, different types of instruments have different sweep
ranges:
DS1204B, <scale_val> range: 1ns/div~50s/div.
DS1104B, <scale_val>range: 2ns/div~50s/div.
DS1074B, <scale_val>range: 5ns/div~50s/div.
In SCAN mode, <scale_val>range: 50ms ~ 50s.