RIGOL Command Systems
Programming Guide for DS1000B Series
:TIMebase[:MAIN]:SCALe <scale_val>
:TIMebase[:MAIN]:SCALe?
Function:
The command sets the timebase scale of main mode, and the unit is s/div.
In NORMAL mode, different types of instruments have different sweep
ranges:
DS1204B , <scale_val> range: 1ns/div~50s/div.
DS1104B, <scale_val>range: 2ns/div~50s/div.
DS1074B, <scale_val>range: 5ns/div~50s/div.
In SCAN mode, <scale_val>range: 50ms ~ 50s.
Returned Format:
The query returns the value of timebase scale, and the unit is s.
Example:
:TIM:MODE MAIN Set the scan mode as main.
:TIM:SCAL 2 Set the timebase scale as 2s.
:TIM:SCAL? Return 2.000e000.
5. :TIMebase:DELayed:SCALe
Command Format:
:TIMebase:DELayed:SCALe <scale_val>
:TIMebase:DELayed:SCALe?
Function:
The command sets the timebase scale of delayed scan, and the unit is s/div.
When the “Delayed” is “ON”, for view waveform details, the waveform may be
amplified under the width of window vary with the delayed timebase scale.
In NORMAL mode, different types of instruments have different sweep
ranges:
DS1204B, <scale_val> range: 1ns/div~50s/div.
DS1104B, <scale_val>range: 2ns/div~50s/div.
DS1074B, <scale_val>range: 5ns/div~50s/div.
In SCAN mode, <scale_val>range: 50ms ~ 50s.

Other manuals for Rigol DS1000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1000B Series and is the answer not in the manual?

Rigol DS1000B Series Specifications

General IconGeneral
BrandRigol
ModelDS1000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals