Command Systems RIGOL
Programming Guide for DS1000B Series
Example:
:TRIG:EDGE:SOUR CHAN1 Set the edge trigger source as channel 1.
:TRIG:EDGE:SOUR? Return CH1.
3. :TRIGger<mode>:LEVel
Command Format:
:TRIGger<mode>:LEVel <level>[,<src>]
:TRIGger<mode>:LEVel? [,<src>]
Function:
The command sets the voltage level of Edge, Pulse or Video trigger.
<mode> may be :EDGE, :PULSe or :VIDEO or :PATTern.
<level> range: (- 6*Scale-Offset①) ~(+6*Scale+Offset①).
Scale is the current vertical scale, and the unit is V/div.
In PATTern mode, <src> should be set to CHANnel<n> or EXT.
NOTE①: Trigger Level range is up to+/-6 Scale, when channel has offset, it needs to detract
offset ,such as 1V tap position, 1V offset, the trigger range is -7V~5V.
Returned Format:
The query returns the value of voltage level, and the unit is V.
Example:
:TRIG:EDGE:LEV 2 Set the trigger level as 2V.
:TRIG:EDGE:LEV? Return 2.000e000.
4. :TRIGger<mode>:SWEep
Command Format:
:TRIGger<mode>:SWEep {AUTO|NORMal|SINGle}
:TRIGger<mode>:SWEep?
Function:
The command sets the trigger mode. The <mode> may be :EDGE, :PULSe
or :PATTern.
AUTO: When no trigger is existing, the system will generate a trigger signal
to force trigger;