EasyManuals Logo

Rigol DS1000B Series Programming Guide

Rigol DS1000B Series
152 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #53 background imageLoading...
Page #53 background image
Command Systems RIGOL
Programming Guide for DS1000B Series
2-41
:TRIG:ALT:TSCAL? SOURB Return 1.000e-003.
5. :TRIGger:ALTernation:TimeOFFSet
Command Format:
:TRIGger:ALTernation:TimeOFFSet <value>[,<source>]
:TRIGger:ALTernation:TimeOFFSet? [<source>]
Function:
The command sets the timebase offset.
In NORMAL mode, <value>: 1s ~ memory capacitance;
In STOP mode, <value: -500s ~ +500s.
Returned Format:
The query returns the value of timebase offset, and the unit is s.
Example:
:TRIG:ALT:TOFFS 0.0002,SOURB Set the timebase offset as 200μs.
:TRIG:ALT:TOFFS? SOURB Return 2.000e-004.
6. :TRIGger:ALTernation:LEVel
Command Format:
:TRIGger:ALTernation:LEVel <value>[,<source>]
:TRIGger:ALTernation:LEVel? [<source>]
Function:
The command sets the trigger level of current channel. The <value> range
<value> range: (- 6*Scale-Offset
) ~(+6*Scale+Offset).Scale is the current
vertical scale, and the unit is V/div. the <source> may be SOURceA or SOURceB,
and the source A and B is different according to the current alternation channel.
NOTE: Trigger Level range is up to+/-6 Scale, when channel has offset, it needs to detract
offset ,such as 1V tap position, 1V offset, the trigger range is -7V~5V.
Returned Format:
The query returns the value of trigger voltage level, and the unit is V.

Other manuals for Rigol DS1000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1000B Series and is the answer not in the manual?

Rigol DS1000B Series Specifications

General IconGeneral
BrandRigol
ModelDS1000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals