Example: 
:TRIG:ALT:VIDEO:MODE ALLLINES,SOURB  Set the sync mode as all lines. 
:TRIG:ALT:VIDEO:MODE? SOURB        Return ALL LINES. 
 
 
13.  :TRIGger:ALTernation:VIDEO:LINE 
Command Format:  
:TRIGger:ALTernation:VIDEO:LINE <value>[,<source>] 
:TRIGger:ALTernation:VIDEO:LINE? [<source>] 
 
Function: 
The command sets the number of sync specified line. In NTSC standard, the 
<value> range is 1~525; in PAL/SECAN standard, the <value> range is 1~625. 
The <source> may be SOURceA or SOURceB, and the source A and B are 
varying with the current alternation channel. 
 
Returned Format:   
The query returns the number of specified line. 
 
Example: 
:TRIG:ALT:VIDEO:LINE 100, SOURB  Set the    specified line number as 100. 
:TRIG:ALT:VIDEO:LINE? SOURB      Return 100. 
 
 
14.  :TRIGger:ALTernation:COUPling 
Command Format: 
:TRIGger:ALTernation:COUPling {DC|AC|LF}[,<source>] 
:TRIGger:ALTernation:COUPling? [<source>] 
 
Function: 
The command sets the coupling mode. 
DC: Allow all signals pass; 
AC: Reject DC signals and attenuate AC signals below 10Hz. 
LF: Reject DC and attenuate low frequency signals below 8kHz. 
The <source> may be SOURceA or SOURceB, and the source A and B are 
varying with the current alternation channel.