EasyManuals Logo

Rigol DS1000B Series Programming Guide

Rigol DS1000B Series
152 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #57 background imageLoading...
Page #57 background image
Command Systems RIGOL
Programming Guide for DS1000B Series
2-45
Example:
:TRIG:ALT:VIDEO:MODE ALLLINES,SOURB Set the sync mode as all lines.
:TRIG:ALT:VIDEO:MODE? SOURB Return ALL LINES.
13. :TRIGger:ALTernation:VIDEO:LINE
Command Format:
:TRIGger:ALTernation:VIDEO:LINE <value>[,<source>]
:TRIGger:ALTernation:VIDEO:LINE? [<source>]
Function:
The command sets the number of sync specified line. In NTSC standard, the
<value> range is 1~525; in PAL/SECAN standard, the <value> range is 1~625.
The <source> may be SOURceA or SOURceB, and the source A and B are
varying with the current alternation channel.
Returned Format:
The query returns the number of specified line.
Example:
:TRIG:ALT:VIDEO:LINE 100, SOURB Set the specified line number as 100.
:TRIG:ALT:VIDEO:LINE? SOURB Return 100.
14. :TRIGger:ALTernation:COUPling
Command Format:
:TRIGger:ALTernation:COUPling {DC|AC|LF}[,<source>]
:TRIGger:ALTernation:COUPling? [<source>]
Function:
The command sets the coupling mode.
DC: Allow all signals pass;
AC: Reject DC signals and attenuate AC signals below 10Hz.
LF: Reject DC and attenuate low frequency signals below 8kHz.
The <source> may be SOURceA or SOURceB, and the source A and B are
varying with the current alternation channel.

Other manuals for Rigol DS1000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1000B Series and is the answer not in the manual?

Rigol DS1000B Series Specifications

General IconGeneral
BrandRigol
ModelDS1000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals