Command Systems RIGOL
Programming Guide for DS1000B Series
2-47
The query returns the value of holdoff time, and the unit is s.
Example:
:TRIG:HOLD 0.0001, SOURA Set the holdoff time of source A as 100us.
:TRIG:HOLD? SOURA Return 1.000e-004.
17. :TRIGger:ALTernation:SENSitivity
Command Format:
:TRIGger:ALTernation:SENSitivity <count>[,<source>]
:TRIGger:ALTernation:SENSitivity? [<source>]
Function:
The command sets the trigger sensitivity of alternation trigger, the count range
is 0.1div~1div. The <source> may be SOURceA or SOURceB, and the source A
and B are varying with to the current alternation channel.
Returned Format:
The query returns the value of trigger sensitivity, and the unit is div.
Example:
:TRIG:ALT:SENS 0.1, SOURceB Set the trigger sensitivity as 01.
:TRIG:ALT:SENS? SOURceB Return 1.000e-001.