EasyManua.ls Logo

Rigol DS1000B Series

Rigol DS1000B Series
152 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RIGOL Command Systems
Programming Guide for DS1000B Series
The command sets the vertical scale for magnifying waveform. The <n> may be
1, 2, 3 or 4.
Probe 0.001X, <range>: 2μV ~ 10mV;
Probe 0.01X, <range>: 20μV ~ 100mV;
Probe 0.1X, <range>: 200μV ~ 1V;
Probe 1X, <range>: 2mV ~ 10V;
Probe 2X, <range>: 4mV ~ 20V;
Probe 5X, <range>: 10mV ~ 50V;
Probe 10X, <range>: 20mV ~ 100V;
Probe 20X, <range>: 40mV ~ 200V;
Probe 50X, <range>: 100mV ~ 500V;
Probe 100X, <range>: 200mV ~ 1kV;
Probe 200X, <range>: 400mV ~ 2kV;
Probe 500X,<range>: 1V ~ 5kV;
Probe 1000X, <range>: 2V ~ 10kV.
Returned Format:
The query returns the value of vertical scale, and the unit is V.
Example:
:CHAN2:PROB 10X Set the attenuation factor of channel 2 as 10X.
:CHAN2:SCAL 20 Set the vertical scale of channel 2 as 20V.
:CHAN2:SCAL? Return 2.000e001.
8. :CHANnel<n>:FILTer
Command Format:
:CHANnel<n>:FILTer {{1|ON}|{0|OFF}}
:CHANnel<n>:FILTer?
Function:
The command sets digital filter function ON or OFF. The <n> may be 1, 2, 3 or
4.
Returned Format:
The query returns 1 or 0, respectively indicates ON or OFF.
Example:

Other manuals for Rigol DS1000B Series

Related product manuals