Command Systems RIGOL
Programming Guide for DS1000B Series
2-59
4. :MEASure:VMIN?
Command Format:
:MEASure:VMIN? [<source>]
Function:
The command measures the minimum of signal from <source>. The <source>
may be CHANnel1, CHANnel2, CHANnel3 or CHANnel4.
Returned Format:
The query returns as -2.640e000, and the unit is V.
5. :MEASure:VAMPlitude?
Command Format:
:MEASure:VAMPlitude? [<source>]
Function:
The command measures the amplitude of signal from <source>. The <source>
may be CHANnel1, CHANnel2, CHANnel3 or CHANnel4.
Returned Format:
The query returns as 5.280e000, and the unit is V.
6. :MEASure:VTOP?
Command Format:
:MEASure:VTOP? [<source>]
Function:
The command measures the top value of signal from <source>. The <source>
may be CHANnel1, CHANnel2, CHANnel3 or CHANnel4.
Returned Format:
The query returns as 2.640e000, and the unit is V.