EasyManuals Logo
Home>Rigol>Test Equipment>DS1000B Series

Rigol DS1000B Series Programming Guide

Rigol DS1000B Series
152 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #76 background imageLoading...
Page #76 background image
RIGOL Command Systems
Programming Guide for DS1000B Series
19. :MEASure:NDUTycycle?
Command Format:
:MEASure:NDUTycycle? [<source>]
Function:
The command measures the negative duty cycle of signal from <source>. The
<source> may be CHANnel1, CHANnel2, CHANnel3 or CHANnel4.
Returned Format:
The query returns as 5.000e001, and the unit is %.
20. :MEASure:PDELay?
Command Format:
:MEASure:PDELay? [<source A>,<source B>]
Function:
The command measures the delay between <sourceA> and <sourceB> relative
to the rising edge. The <source> may be CHANnel1, CHANnel2, CHANnel3 or
CHANnel4.
Returned Format:
The query returns as <-1.000 e-004, and the unit is s.
21. :MEASure:NDELay?
Command Format:
:MEASure:NDELay? [<source A>,<source B>]
Function:
The command measures the delay between <sourceA> and <sourceB> relative
to the falling edge. The <source> may be CHANnel1, CHANnel2, CHANnel3 or
CHANnel4.
Returned Format:
The query returns as <-1.000 e-004, and the unit is s.

Other manuals for Rigol DS1000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1000B Series and is the answer not in the manual?

Rigol DS1000B Series Specifications

General IconGeneral
BrandRigol
ModelDS1000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals