EasyManuals Logo

Rigol DS1000B Series Programming Guide

Rigol DS1000B Series
152 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #82 background imageLoading...
Page #82 background image
RIGOL Command Systems
Programming Guide for DS1000B Series
1. :WAVeform:FORMat
Command Format:
:WAVeform:FORMat <value>
:WAVeform:FORMat?
Function:
The command sets the format of waveform data. The <value> may be WORD,
BYTE or ASCii.
Difference of WORD, BYTE and ASCii:
ASCii: Returns ASCII values when data ware transformed into character.
Eg: Waveform data is 1000, returns ’1’,’0’,’0’,’0’, a point correspond many bytes.
BYTE and WORD: They will returns 8 bit and 16 bit values to use datum.
Eg: Waveform data is 1000, returns 1000 in decimal system, a point correspond
one byte.
Returned Format:
The query returns WORD or BYTE, ASCii.
Example:
:WAV:FORM ASC Set the data format as ASCII.
:WAV:FORM? Return ASCii.
2. :WAVeform:DATA?
Command Format:
:WAVeform:DATA? [<source>]
Function:
The command reads waveform data from the specified source. <source> may
be : CHANnel1, CHANnel2, CHANnel3, CHANnel4 or MATH.
Returned Format:
The query returns a certain amount of waveform data that specifed
by :WAVeform:POINts.
NOTE:
The command returns the data on the screen while waveform playback, at this moment,

Other manuals for Rigol DS1000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1000B Series and is the answer not in the manual?

Rigol DS1000B Series Specifications

General IconGeneral
BrandRigol
ModelDS1000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals