Loading...
RIGOL Command Systems
Programming Guide for DS1000B Series
1. :WAVeform:FORMat
Command Format:
:WAVeform:FORMat <value>
:WAVeform:FORMat?
Function:
The command sets the format of waveform data. The <value> may be WORD,
BYTE or ASCii.
Difference of WORD, BYTE and ASCii:
ASCii: Returns ASCII values when data ware transformed into character.
Eg: Waveform data is 1000, returns ’1’,’0’,’0’,’0’, a point correspond many bytes.
BYTE and WORD: They will returns 8 bit and 16 bit values to use datum.
Eg: Waveform data is 1000, returns 1000 in decimal system, a point correspond
one byte.
Returned Format:
The query returns WORD or BYTE, ASCii.
Example:
:WAV:FORM ASC Set the data format as ASCII.
:WAV:FORM? Return ASCii.
2. :WAVeform:DATA?
Command Format:
:WAVeform:DATA? [<source>]
Function:
The command reads waveform data from the specified source. <source> may
be : CHANnel1, CHANnel2, CHANnel3, CHANnel4 or MATH.
Returned Format:
The query returns a certain amount of waveform data that specifed
by :WAVeform:POINts.
NOTE:
The command returns the data on the screen while waveform playback, at this moment,

Other manuals for Rigol DS1000B Series