RIGOL Chapter 3 Performance Test
Service Guide for DS1000E, DS1000D Series
3-10
Figure 3-12 Setting interface
7. Press LA button on the oscilloscope, select D7-D0 and D15-D8 to open all the
logic channels.
8. Select EDGE mode under the trigger menu and select any one source among Do
to D15, if obtain the stable interface as follows, the LA module will be proved
works normally.
Figure 3-13 Edge trigger of digital channel